FMS Image Data Processing

 

 

ODP logging contractor: LDEO-BRG

Hole: 1253A

Leg: 205

Location: Costa Rica Convergent Margin (equatorial NE Pacific)

Latitude: 9° 38.8583' N

Longitude: 86° 11.4337' W

Logging date: October 7-8, 2002

Bottom felt: 4387 mbrf

Total penetration: 600 mbsf

Total core recovered:  148.6 m (64.6 %)

 

FMS Pass 1: 410 - 528 mbsf

FMS Pass 2: 410 - 560 mbsf

Magnetic declination: 1.393°

Water depth: 4390 mbrf

 

The basic principle of the FMS (Formation MicroScanner) is to map the conductivity of the borehole wall with a dense array of sensors. This provides a high resolution electrical image of the formation which can be displayed in either gray or color scale. The purpose of this report is to describe the images from Leg 205 and the different steps used to generate them from the raw FMS measurements.

 

The FMS tool records 4 perpendicular electrical images, using four pads, which are pressed against the borehole wall. Each pad has 16 buttons and the tool provides approximately 25% coverage of the borehole wall. The tool string also contains a triaxial accelerometer and three flux-gate magnetometers (in the GPIT, General Purpose Inclinometry Tool) whose results are used to accurately orient and position the images. Measurements of hole size, cable speed, and natural gamma ray intensity also contribute to the processing.

 

Data Quality

 

Hole conditions were variable because of the contrasting lithologies  encountered. The hole was wider than 18 inches in the sediment section (429-459 mbsf), with at least two of the pads having little contact with the borehole wall.  The hole was in excellent condition in the gabbro (414-428 mbsf and 458-558 mbsf), the diameter being mostly 12 inches, except for isolated rougher sections  where the diameter could reach 18 inches.  Consequently, the FMS images are generally of very good quality in the gabbro. The hole was cased to 414 mbsf.

 

In Pass 2, the tool string stuck in three places, leading to localized depth discrepancies  (around 468, 475, 542-546 mbsf).  In Pass 1 at 486 and 491 mbsf, the high resistivity bands in the image are artifacts. Only one of the caliper  logs (C1) recorded properly for both passes. Although the C2 caliper was open, as is seen from the good contact in the images, it did not record good measurements.

 

Image Processing

 

Processing is required to convert the electrical current in the formation, emitted by the FMS button electrodes, into a gray or color-scale image representative of the conductivity changes. This is achieved through two main processing phases: data restoration and image display.

 

1) Data Restoration

 

Speed Correction. The data from the z-axis accelerometer is used to correct the vertical position of the data for variations in the speed of the tool (GPIT speed correction), including stick and slip. In addition, image-based speed correction is also applied to the data: the principle behind this is that if the GPIT speed correction is successful, the readings from the two rows of buttons on the pads will line up, and if not, they will be offset from each other (a zigzag effect on the image).

 

Equalization: Equalization is the process whereby the average response of all the buttons of the tool are rendered approximately the same over large intervals, to correct for various tool and borehole effects which affect individual buttons differently. These effects include differences in the gain and offset of the pre-amplification circuits associated with each button, and differences in contact with the borehole wall between buttons on a pad, and between pads.

 

Button Correction. If the measurements from a button are unreasonably different from its neighbors (e.g. dead buttons) over a particular interval, they are declared faulty, and the defective trace is replaced by traces from adjacent good buttons.

 

EMEX voltage correction. The button response (current) is controlled by the EMEX voltage, which is applied between the button electrode and the return electrode. The EMEX voltage is regulated to keep the current response within the operating range. The button response is divided by the EMEX voltage so that the response corresponds more closely to the conductivity of the formation.

 

Depth-shifting: Each of the logging runs are depth-matched to a common scale by means of lining up distinctive features of the natural gamma log from each of the tool strings. If the reference logging run is not the FMS tool string, the specified depth shifts are applied to the FMS images. The position of data located between picks is computed by linear interpolation.

 

2) Image Display:

In "static normalization", a histogram equalization technique is used to obtain the maximum quality image. In this technique, the resistivity range of the entire interval of good data is computed and partitioned into 256 color levels. This type of normalization is best suited for large-scale resistivity variations.

 

The image can be enhanced when it is desirable to highlight features in sections of the well where resistivity events are relatively subdued when compared with the overall resistivity range in the section. This enhancement is called "dynamic normalization". By rescaling the color intensity over a smaller interval, the contrast between adjacent resistivity levels is enhanced. It is important to note that with dynamic normalization, resistivities in two distant sections of the hole cannot be directly compared with each other. A 2-m normalization interval is used.

 

Oriented Presentation: The image is displayed as an unwrapped borehole cylinder (its circumference is derived from the bit size). Several passes can be oriented and merged together on the same presentation to give additional borehole coverage if the tool pads followed a different track. A dipping plane in the borehole will be displayed as a sinusoid on the image; the amplitude of this sinusoid is proportional to the dip of the plane. The images are oriented with respect to north; hence the strike of dipping features can also be determined.

 